Research Careers Contact

GlossaryHipot test

Hipot test

A hipot test, short for high-potential test and also called a dielectric withstand test, applies a voltage much higher than normal operating voltage across a product's insulation for a set time and checks that the insulation does not break down. It proves that isolation barriers, such as between a high-voltage battery and user-accessible parts, hold.

How it works

The tester connects one output to the circuits on one side of the insulation, often all high-voltage conductors tied together, and the other to the other side, such as the chassis or low-voltage circuits. It ramps the voltage up, holds it, and ramps down, while measuring leakage current. If current exceeds a set limit or an arc occurs, the unit fails.

Test voltage, duration and AC or DC choice come from the product safety standard that applies, based on working voltage, insulation type and overvoltage category. IEC 61010-1 for measurement and control equipment and similar product standards set these values, and IEC 60664-1 covers insulation coordination, the principles behind clearance, creepage and test voltages. Type testing on design samples uses the full values. Production lines run shorter routine tests on every unit.

Hipot test vs insulation resistance test

An insulation resistance test applies a moderate DC voltage and measures resistance, usually in megohms, to show insulation condition. A hipot test applies a stress voltage well above operating levels to show the insulation withstands overvoltage without breakdown. The first measures quality, the second proves strength. Many production lines run both.

Why it matters

Robots and drones with high-voltage batteries, motor drives or mains chargers rely on insulation to keep people safe and keep high voltage out of low-voltage electronics. Defects such as a pinched wire, contamination on the board, a cracked isolator or insufficient creepage are invisible to functional tests. A hipot test finds them before the product ships, and safety certification requires it.

How to get it right

  • Take test voltage, duration and limits from the applicable safety standard, not from habit.
  • Design creepage and clearance distances and isolation components for the test voltage from the start.
  • Tie together all conductors on each side, and protect sensitive circuits that are not part of the test.
  • Use a tester with ramp control, arc detection and secure fixturing for operator safety.
  • Log results for every unit, with serial numbers, for traceability.

Hipot tests in robotics and drone hardware

Robots with high-voltage battery packs and motor drives use galvanic isolation between power and control electronics, and hipot tests confirm it in production. Chargers and docks that connect to mains need routine hipot testing as part of their safety certification. Board designs carry the creepage slots, isolation components and test access that make the test meaningful.

Learn more